CVE-2024-48966
CVE-2024-48966
10.0
CriticalPublished:
Last updated:
Source:productsecurity@baxter.com
Deferred
Weakness (CWE)
CVSS Vector
v3.1- Attack Vector
- Network
- Attack Complexity
- Low
- Privileges Required
- None
- User Interaction
- None
- Scope
- Changed
- Confidentiality
- High
- Integrity
- High
- Availability
- High
Description
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
References
productsecurity@baxter.com
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01