CVE-2024-48973
CVE-2024-48973
9.3
CriticalPublished:
Last updated:
Source:productsecurity@baxter.com
Deferred
Weakness (CWE)
CVSS Vector
v3.1- Attack Vector
- Local
- Attack Complexity
- Low
- Privileges Required
- None
- User Interaction
- None
- Scope
- Changed
- Confidentiality
- High
- Integrity
- High
- Availability
- High
Description
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
References
productsecurity@baxter.com
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01