CVE-2024-9832
CVE-2024-9832
9.3
CriticalPublished:
Last updated:
Source:productsecurity@baxter.com
Deferred
Weakness (CWE)
CVSS Vector
v3.1- Attack Vector
- Local
- Attack Complexity
- Low
- Privileges Required
- None
- User Interaction
- None
- Scope
- Changed
- Confidentiality
- High
- Integrity
- High
- Availability
- High
Description
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
References
productsecurity@baxter.com
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01